Simple compact system for the extraction of radon from water samples
Journal Article
·
· Am. Ind. Hyg. Assoc. J.; (United States)
This paper describes a compact system for the extraction of radon from water samples for subsequent quantitation. The system can accommodate samples having different volumes. The volumes tested ranged from 130 to 455 mL. The limit of detection of the 155 mL sample, based upon a counting time of 1000 min, is 0.034 Bq/L (0.91 pCi/L) of water.
- Research Organization:
- Manitoba Hydro, Winnipeg
- OSTI ID:
- 6529817
- Journal Information:
- Am. Ind. Hyg. Assoc. J.; (United States), Journal Name: Am. Ind. Hyg. Assoc. J.; (United States) Vol. 48:3; ISSN AIHAA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Determination of radon-222 in ground water using liquid scintillation counting-survey of carefree cave-creek water basin in Arizona
Variation of radon sources and indoor radon along the southwestern edge of the Canadian shield
Determination of radium in water
Conference
·
Wed Dec 30 23:00:00 EST 1992
·
OSTI ID:45172
Variation of radon sources and indoor radon along the southwestern edge of the Canadian shield
Journal Article
·
Sat Dec 31 23:00:00 EST 1988
· Environment International; (USA)
·
OSTI ID:6613980
Determination of radium in water
Technical Report
·
Sun Oct 01 00:00:00 EDT 1995
·
OSTI ID:148676