Performance testing of beta dosimeters used at Department of Energy facilities

Conference ·
OSTI ID:6528647
A performance test based on the American national draft standard N13.11 was conducted for dosimeter systems in use at Department of Energy facilities. The large differences in dosimeter response found were due to use of different calibration source standards and different dosimeter designs. Differences in /sup 90/Sr//sup 90/Y calibrations were approximately 20% or less for all but one participant. The differences observed were attributed to variable thicknesses of dosimeter elements and variable source irradiation geometries. Improved beta calibration standards will result if irradiation specifications include acceptable ranges from the depth-dose characteristics. The low-energy beta responses observed were consistent with the thicknesses of dosimeter sensitive elements and overlying filtration.