Radometer; A portable field instrument for the rapid measurement of environmental radon and thoron

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA)
OSTI ID:6487258
 [1]
  1. Environmental Measurements Lab., US DOE, New York, NY (US)

This paper reports on the Radometer, which measures radon and thoron by electrostatic collection of charged progeny on a surface barrier detector. Accurate measurements are made in only a few minutes because this instrument uses a dual electric field configuration, thus eliminating filters which slow the response time. An on board computer corrects for the effect of moisture and previous deposition to maintain accuracy and fast response when the instrument encounters high humidity and/or rapid changes in concentration. The instrument's detection volume is 6 L and its efficiency is 25% or .041 cpm per Bq m{sup {minus}3} (1.5 cpm per pCi L{sup {minus}1}).

OSTI ID:
6487258
Report Number(s):
CONF-900143--
Journal Information:
IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA), Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) Vol. 37:2; ISSN 0018-9499; ISSN IETNA
Country of Publication:
United States
Language:
English