Using a digital signal processor as a data stream controller in digital subtraction angiography

Journal Article · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
DOI:https://doi.org/10.1109/23.212337· OSTI ID:6433244
;  [1]
  1. Univ. of California, Berkeley (United States)
High speed, flexibility, and good arithmetic abilities make digital signal processors (DSP) a good choice as input/output controllers for real time applications. The DSP can be made to pre-process data in real time to reduce data volume, to open early windows on what is being acquired and to implement input/output controller for a digital subtraction angiographic imaging system. The DSP pre-processes the raw data, reducing data volume by a factor of two, and is potentially capable of producing real-time subtracted images for immediate display.