Analysis of peroxyacyl nitrates by gas chromatography with electron capture detection

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac60197a028· OSTI ID:6411288

A method is described for the measurement of concentrations of peroxyacetyl nitrate (PAN) and peroxypropionyl nitrate (PPN) from polluted air samples. The electron capture detector offers an extremely sensitive and selective method of measuring peroxyacyl nitrates in the parts per hundred million range using untreated air samples of less than 5 ml. Further work on the calibration factors for the various homolog remains to be done with studies of the transportation of samples to the laboratory. The results to date demonstrate that the peroxyacyl nitrates can be measured at less than 1 pphm with the electron capture detector using very small untreated samples.