Automatic sample changers maintenance manual
This manual describes and provides trouble-shooting aids for the Automatic Sample Changer electronics on the automatic beta counting system, developed by the Los Alamos Scientific Laboratory Group CNC-11. The output of a gas detector is shaped by a preamplifier, then is coupled to an amplifier. Amplifier output is discriminated and is the input to a scaler. An identification number is associated with each sample. At a predetermined count length, the identification number, scaler data plus other information is punched out on a data card. The next sample to be counted is automatically selected. The beta counter uses the same electronics as the prior count did, the only difference being the sample identification number and sample itself. This manual is intended as a step-by-step aid in trouble-shooting the electronics associated with positioning the sample, counting the sample, and getting the needed data punched on an 80-column data card.
- Research Organization:
- Los Alamos Scientific Lab., NM (USA)
- DOE Contract Number:
- W-7405-ENG-36
- OSTI ID:
- 6396279
- Report Number(s):
- LA-6759-M
- Country of Publication:
- United States
- Language:
- English