Single-laboratory evaluation and modification of US EPA (Environmental Protection Agency) Methods 7470 and 7471 for the determination of mercury in aqueous and solid hazardous wastes (journal version)
The current U.S. Environmental Protection Agency (U.S. EPA) protocols for mercury determinations in aqueous and solid-waste samples (SW-846 Methods 7470 and 7471) using recirculating cold-vapor atomic absorption spectrometry (CV-AAS) were evaluated. The U.S. EPA protocols have been modified in a single-laboratory study to facilitate additional QC measures, to enhance detectability for low-level mercury concentrations, and to eliminate nonspecific vapor-absorption interferences. Volumetric manipulations for additional QC measures, if required, are facilitated by performing the sample digestions in Erlenmeyer flasks rather than in the current Biochemical Oxygen Demand (BOD) reduction-aeration bottles. Instrument detectability is improved 10-fold by using a gas sparging bottle as a dedicated reduction-aeration vessel and a silver wool-amalgamation CV-AAS system operated in an open configuration. The on-line amalgamation thermal-desorption process of the modified CV-AAS system eliminates interfering water and organic matrix vapors prior to the mercury absorption measurement. Good accuracy and precision have been obtained with the amalgamation CV-AAS system for the analyses of four reference sediment materials.
- Research Organization:
- Battelle Columbus Div., OH (USA)
- OSTI ID:
- 6368001
- Report Number(s):
- PB-89-130389/XAB
- Resource Relation:
- Other Information: Pub. in Environment International, Vol. 13, 475-489(1987)
- Country of Publication:
- United States
- Language:
- English
Similar Records
Interlaboratory evaluation of SW-846 methods 7470 and 7471 for the determination of mercury in environmental samples
Mercury Continuous Emmission Monitor Calibration