Title: Design and evaluation of HEADTOME-IV, a whole-body positron emission tomograph

Conference · · IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:6359461

A whole body positron emission tomograph HEADTOME-IV has been developed, and its physical performances were investigated. The in-plane spatial resolution of 4.5 mm was realized with stationary-sampling at the center of the field-of-view. The axial slice thickness was 9.5 and 9.0-mm for direct and cross planes, respectively. By moving the gantry framework axially, transaxial images of 14 or 21 slices are obtained quasi-simultaneously. The realtime-operation large-scale cache memory system was effective to realize realtime corrections for deadtime and radionuclide decay, and realtime weighted integration for the purpose of a rapid calculation of rate-constant images.

Research Organization:
Research Institute for brain and blood vessels-Akita, Senshuu-Kubota-Machi, Akita City, Akita 010 (JP)
OSTI ID:
6359461
Report Number(s):
CONF-881103-; TRN: 89-009837
Journal Information:
IEEE Trans. Nucl. Sci.; (United States), Vol. 36:1; Conference: IEEE nuclear science symposium, Orlando, FL, USA, 9 Nov 1988
Country of Publication:
United States
Language:
English