Gamma-ray spectrometric determination of UF/sub 6/ assay with 1 percent precision for international safeguards. Part 1: product and feed in 1S and 2S sample cylinders

Technical Report ·
DOI:https://doi.org/10.2172/6348245· OSTI ID:6348245
The method is based on counting the 186-keV gamma rays emitted by /sup 235/U using a Pb-collimated Ge(Li) detector. Measurements of fifty UF/sub 6/ product and feed cylinders reveal the following precisions and counting times: Product - 2S, 0.98% (600 s); Feed - 2S, 0.48% (2500 s); Product - 1S, 0.62% (1000 s); Feed - 1S, 0.73% (3000 s). A 1% precision is desired for variables - attributes verification measurements of /sup 235/U assay in UF/sub 6/ sample cylinders for safeguards inspections by the International Atomic Energy Agency (IAEA). Statistically, these measurements stand between fine, high-precision (or variables) measurements and gross, low-precision (or attributes) ones. Because of their intermediate precisions, the variables-attributes measurements may not require analysis of all samples, and this could result in significant savings of IAEA inspector time. Although the precision of the above results is satisfactory, the average relative differences between gamma-ray and mass-spectrometric determinations for the last two sets of measurements (1S cylinders) have positive biases.