Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Shielding requirements for constant-potential diagnostic x-ray beams determined by a Monte Carlo calculation

Journal Article · · Health Phys.; (United States)
A Monte Carlo calculation has been performed to determine the transmission of broad constant-potential x-ray beams through Pb, concrete, gypsum wallboard, steel and plate glass. The EGS4 code system was used with a simple broad-beam geometric model to generate exposure transmission curves for published 70, 100, 120 and 140-kVcp x-ray spectra. These curves are compared to measured three-phase generated x-ray transmission data in the literature and found to be reasonable. For calculation ease the data are fit to an equation previously shown to describe such curves quite well. These calculated transmission data are then used to create three-phase shielding tables for Pb and concrete, as well as other materials not available in Report No. 49 of the NCRP.
Research Organization:
St. Luke's Medical Center, Milwaukee, WI (USA)
OSTI ID:
6326016
Journal Information:
Health Phys.; (United States), Journal Name: Health Phys.; (United States) Vol. 56:2; ISSN HLTPA
Country of Publication:
United States
Language:
English