X-ray gauge measures areal density variations as small as 0. 1%
A gauge measures areal density variations as small as 0.1% in low density plastic foams and other materials. It has a spatial resolution of 0.5 mm. This system is safe, portable, easy to use, and is readily automated. The 6 keV x-ray flux from a /sup 55/Fe source is collimated to a 0.5 mm diameter beam and attenuated by the sample. Transmitted x-rays are detected with a proportional gas detector and counted with a scaler-timer. After measuring the x-ray attenuation in two areas of the sample, the areal density variation is calculated using an