Determining yttrium in plutonium by anion-exchange x-ray fluorescence
This report describes a method for determining yttrium in plutonium using an anion-exchange separation and x-ray fluorescence. We add zirconium to the plutonium solution as an internal standard. We oxidize the plutonium to Pu + 4 and pass the solution through an anion-exchange column with 8M HCl. The Pu + 4 sorbs to the resin and the yttrium and zirconium pass through completely. We evaporate the eluate solution containing the yttrium and zirconium and transfer it to a 10-ml volumetric flask. We add a portion of this solution to an x-ray cell and measure the Ka x-ray line for both yttrium and zirconium. The ratio of yttrium to zirconium is then compared with standards. This method has a precision of 0.84% relative standard deviation for yttrium over a concentration range of 0.5 to 3.5 mg in a 10-ml volume. 1 ref., 2 figs., 2 tabs.
- Research Organization:
- Los Alamos National Lab., NM (USA)
- DOE Contract Number:
- W-7405-ENG-36
- OSTI ID:
- 6302516
- Report Number(s):
- LA-10544; ON: DE86005840
- Country of Publication:
- United States
- Language:
- English
Similar Records
Anion exchange x-ray-fluorescence determination of uranium in plutonium
Determining gallium from plutonium using anion exchange and x-ray fluorescence
Related Subjects
400103* -- Radiometric & Radiochemical Procedures-- (-1987)
400105 -- Separation Procedures
ACCURACY
ACTINIDES
CALIBRATION STANDARDS
CHEMICAL ANALYSIS
DATA
ELEMENTS
EXPERIMENTAL DATA
INFORMATION
ION EXCHANGE
METALS
NONDESTRUCTIVE ANALYSIS
NUMERICAL DATA
PLUTONIUM
QUANTITATIVE CHEMICAL ANALYSIS
QUANTITY RATIO
RELIABILITY
STANDARDS
TRANSITION ELEMENTS
TRANSURANIUM ELEMENTS
X-RAY EMISSION ANALYSIS
X-RAY FLUORESCENCE ANALYSIS
YTTRIUM
ZIRCONIUM