Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

TFTR horizontal high-resolution Bragg x-ray spectrometer

Technical Report ·
DOI:https://doi.org/10.2172/6296761· OSTI ID:6296761

A bent quartz crystal spectrometer of the Johann type with a spectral resolution of lambda/..delta..lambda = 10,000 to 25,000 is used on TFTR to determine central plasma parameters from the spectra of heliumlike and lithiumlike metal impurity ions (Ti, Cr, Fe, and Ni). The spectra are observed along a central radial chord and are recorded by a position sensitive multiwire proportional counter with a spatial resolution of 250. Standard delay-line time-difference readout is employed. The data are histogrammed and stored in 64k of memory providing 128 time groups of 512-channel spectra. The central ion temperature and the toroidal plasma rotation are inferred from the Doppler broadening and Doppler shift of the K lines. The central electron temperature, the distribution of ionization states, and dielectronic recombination rates are obtained from satellite-to-resonance line ratios. The performance of the spectrometer is demonstrated by measurements of the Ti XXI K radiation.

Research Organization:
Princeton Univ., NJ (USA). Plasma Physics Lab.
DOE Contract Number:
AC02-76CH03073
OSTI ID:
6296761
Report Number(s):
PPPL-2173; ON: DE85004397
Country of Publication:
United States
Language:
English

Similar Records

Tokamak Fusion Test Reactor horizontal high-resolution Bragg x-ray spectrometer (abstract)
Journal Article · Sat Jun 01 00:00:00 EDT 1985 · Rev. Sci. Instrum.; (United States) · OSTI ID:5761449

Tokamak Fusion Test Reactor horizontal high-resolution Bragg x-ray spectrometer (abstract)
Journal Article · Wed May 01 00:00:00 EDT 1985 · Rev. Sci. Instrum.; (United States) · OSTI ID:5545601

Vertical high-resolution Bragg x-ray spectrometer for the tokamak fusion test reactor
Journal Article · Fri Aug 01 00:00:00 EDT 1986 · Rev. Sci. Instrum.; (United States) · OSTI ID:5564980