Calibration of moisture monitors
A method for calibrating an aluminum oxide hygrometer against an optical chilled mirror dew-point hygrometer has been established. A theoretical cross-point line of dew points from both hygrometers and a maximum moisture content of 10 ppM/sub v/ are used to define an area for calibrating the sensor probes of the aluminum oxide hygrometer.
- Research Organization:
- Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
- DOE Contract Number:
- W-7405-ENG-36
- OSTI ID:
- 6279948
- Report Number(s):
- LA-7696-MS; TRN: 79-009264
- Country of Publication:
- United States
- Language:
- English
Similar Records
Comparison of continuous moisture monitors in the range 1 to 15 ppm
Viking Lander: subsurface water analyzing probe. [Mars subsoil]
Measure moisture in a vapor
Journal Article
·
Tue Oct 01 00:00:00 EDT 1985
· Rev. Sci. Instrum.; (United States)
·
OSTI ID:6279948
Viking Lander: subsurface water analyzing probe. [Mars subsoil]
Technical Report
·
Wed Oct 01 00:00:00 EDT 1969
·
OSTI ID:6279948
Measure moisture in a vapor
Journal Article
·
Thu Nov 01 00:00:00 EST 1973
· Hydrocarbon Process.; (United States)
·
OSTI ID:6279948
Related Subjects
42 ENGINEERING
47 OTHER INSTRUMENTATION
GLOVEBOXES
MOISTURE GAGES
CALIBRATION
ALUMINIUM OXIDES
ALUMINIUM COMPOUNDS
CHALCOGENIDES
LABORATORY EQUIPMENT
MEASURING INSTRUMENTS
OXIDES
OXYGEN COMPOUNDS
420202* - Engineering- Protective Structures & Equipment
440300 - Miscellaneous Instruments- (-1989)
47 OTHER INSTRUMENTATION
GLOVEBOXES
MOISTURE GAGES
CALIBRATION
ALUMINIUM OXIDES
ALUMINIUM COMPOUNDS
CHALCOGENIDES
LABORATORY EQUIPMENT
MEASURING INSTRUMENTS
OXIDES
OXYGEN COMPOUNDS
420202* - Engineering- Protective Structures & Equipment
440300 - Miscellaneous Instruments- (-1989)