Radiation protection aspects of a new high-energy linear accelerator
Journal Article
·
· Med. Phys.; (United States)
The Therac-25 is a new 25-MeV linear accelerator manufactured by Atomic Energy of Canada, Ltd. The first two units have recently been installed in Toronto, Ontario and Halifax, Nova Scotia. Calculations and measurements of primary and secondary radiation levels were made. Neutron dose-equivalent rates were measured inside and outside the room. The maximum leakage rate at 1 m from the accelerator target was 0.4% Sv per peak photon Gy. The tenth value layer for neutrons from the Therac-25, at the entrance to a one-legged maze was found to be 5.5 cm of polyethylene. Measurements were done to estimate daily technologist exposure due to induced activity in the treatment room.
- Research Organization:
- Division of Clinical Physics, Toronto-Bayview Clinic, Ontario Cancer Foundation, Toronto, Ontario M4N 3M5, Canada
- OSTI ID:
- 6246538
- Journal Information:
- Med. Phys.; (United States), Vol. 12:1
- Country of Publication:
- United States
- Language:
- English
Similar Records
Neutron dose and energy spectrum outside a 20-MV accelerator treatment room
The neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room
First production off Canada to start in '92
Journal Article
·
Mon Sep 01 00:00:00 EDT 1986
· Med. Phys.; (United States)
·
OSTI ID:6246538
+2 more
The neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room
Journal Article
·
Tue Mar 15 00:00:00 EDT 2011
· Medical Physics
·
OSTI ID:6246538
+3 more
First production off Canada to start in '92
Journal Article
·
Wed Mar 06 00:00:00 EST 1991
· Oil and Gas Journal; (United States)
·
OSTI ID:6246538
+1 more