Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Influence of carrier strontium level and eluant volume on the performance of (/sup 82/Sr)--(/sup 82/Rb) biomedical generators

Journal Article · · J. Nucl. Med.; (United States)
OSTI ID:6209767
Previous inconsistencies in the medical literature regarding the performance characteristics of (Sr-82)--(Rb-82) radionuclide generators have been partially resolved by further experimentation. The Chelex generator is dependent upon the presence of carrier strontium, and degrades appreciably below 0.5-mg column loadings. The Bio-Rex generator is not similarly affected, however. Chelex also exhibits a ''self-cleaning'' effect after initial loading, and this resin withstands generator deterioration produced by repetitive elution better than does Bio-Rex.
Research Organization:
Los Alamos Scientific Lab., NM
OSTI ID:
6209767
Journal Information:
J. Nucl. Med.; (United States), Journal Name: J. Nucl. Med.; (United States) Vol. 19:11; ISSN JNMEA