Title: Application of a semi-empirical Ge(Li) detector calibration technique for large-volume samples

Technical Report ·
OSTI ID:6196892

A previously developed semi-empirical method for the calculation of the absolute peak efficiency of coaxial Ge(Li) detectors for large volume samples is reviewed and applied using a computer model and experimental determinations of calibration parameters. The efficiencies calculated for extended air, lucite and aluminum source materials are compared with those obtained with spiked sample calibrations and point source calibrations. The errors are within +-5% in most cases. For the calibrations, two versions of computer programs written in FORTRAN are presented.

Research Organization:
USDOE Environmental Measurements Lab., New York
OSTI ID:
6196892
Report Number(s):
DOE/EML-409; ON: DE83012801
Resource Relation:
Other Information: Portions are illegible in microfiche products
Country of Publication:
United States
Language:
English