Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Gas chromatographic determination of volatile sulfides at trace levels in natural fresh waters

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00177a007· OSTI ID:6190795
Volatile sulfides in the aqueous phase were analyzed by degassing the solution and cryogenically trapping the volatiles. A gas chromatograph equipped with a Hall electrolytic conductivity detector is used in the sulfur mode. Water samples were analyzed with the internal standard method because of variable recoveries. Recovery efficiencies of the internal standard show a good linear relationship with the analytes. Detection limits (as S) of 0.03 ng/L for H/sub 2/S, 0.02 ng/L for COS, 0.06 ng/L for CH/sub 3/SH, 0.09 ng/L for CS/sub 2/, 0.05 ng/L for CH/sub 3/SCH/sub 3/, 0.07 ng/L for CH/sub 3/CH/sub 2/CH/sub 2/SH, and 0.10 ng/L for CH/sub 3/SSCH/sub 3/ can be achieved for a 100-mL sample. The precision is better than 10% for the range of 1-4000 ng of S/L.
Research Organization:
McMaster Univ., Hamilton, Ontario (Canada)
OSTI ID:
6190795
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 61:2; ISSN ANCHA
Country of Publication:
United States
Language:
English