Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Magnification mammography: a low-dose technique

Conference · · Radiology; (United States)
OSTI ID:6187203

Image quality and radiation exposures of a mammographic technique using direct radiographic magnification at 2 X with a microfocal spot x-ray tube and a fast, double screen-film system were compared to those of conventional contact mammography with a rare-earth screen and molybdenum target tube. The results indicate that the magnifiction technique yields improved detection of microcalcifications and comparable visualization of soft-tissue details, with a large reduction in radiation exposure. This technique has demonstrated the feasibility of carrying out high-quality mammography with an entrance dose of 1.35 x 10/sup -3/ Gy (135 mrad) for the average breast.

OSTI ID:
6187203
Journal Information:
Radiology; (United States), Journal Name: Radiology; (United States) Vol. 131:3; ISSN RADLA
Country of Publication:
United States
Language:
English