Two-dimensional filtering of SPECT images using the Metz and Wiener filters

Journal Article · · J. Nucl. Med.; (United States)
OSTI ID:6153507

Two-dimensional filtering, both before and after reconstruction, has been applied to the processing of single photon emission computerized tomographic (SPECT) images. The filters investigated were the count-dependent Metz filter and Wiener filter, both of which automatically adapt to the image being processed. Using a SPECT phantom, with images reconstructed with these filters rather than the ramp, the authors observed a statistically signficant increase in the image contrast for solid Plexiglas spheres, and significant decrease in the percent fractional standard deviation of counts in a region of uniform activity. The adaptability of these filters is demonstrated by a comparison of SPECT acquisition of the phantom at two different count levels. An example of the application to clinical studies is presented. Two-dimensional digital image restoration with these techniques can produce a significant increase in SPECT image quality, with a small cost in processing time when these techniques are implemented on an array processor.

Research Organization:
Univ. of Massachusetts Medical Center, Worcester
OSTI ID:
6153507
Journal Information:
J. Nucl. Med.; (United States), Journal Name: J. Nucl. Med.; (United States) Vol. 25:11; ISSN JNMEA
Country of Publication:
United States
Language:
English