Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Improvement of small-field penumbra and dose distributions on a 4 MV accelerator

Journal Article · · Int. J. Radiat. Oncol., Biol. Phys.; (United States)
The penumbra and dose distributions for small x ray fields from a 4 MV accelerator, designed for 80 cm source-to-surface (SSD) but used at 100 cm SSD, were observed to be similar to those obtained from cobalt-60 units. Penumbra was reduced and flatter isodose curves were obtained by the addition of tungsten trimmers at a distance of 48 cm from the x ray source. This stratagem improves the definition of irradiation volume of small fields which are common in the treatment of lesions in the head and neck.
Research Organization:
Yale Univ., New Haven, CT
OSTI ID:
6143155
Journal Information:
Int. J. Radiat. Oncol., Biol. Phys.; (United States), Journal Name: Int. J. Radiat. Oncol., Biol. Phys.; (United States) Vol. 7:7; ISSN IOBPD
Country of Publication:
United States
Language:
English