Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

A monitor for the separate determination of HT and HTO

Journal Article · · IEEE Trans. Nucl. Sci.; (United States)
A prototype monitor that discriminates between tritiated water (HTO) and tritiated hydrogen (HT) in air is discussed. The separate determination of these two species is based upon their different permeation rates through the polymer Nafion (DuPont). A single diffusion stage using this material is sufficient to reduce the concentration of HT relative to HTO by a factor of 10/sup 3/ while still maintaining an acceptable time response. The HTO contribution to the measured HT concentration is also small. The electronic and mechanical design of this monitor is outlined and some preliminary test results presented.
Research Organization:
Chalk River Nuclear Lab., Ontario, Canada
OSTI ID:
6132089
Journal Information:
IEEE Trans. Nucl. Sci.; (United States), Journal Name: IEEE Trans. Nucl. Sci.; (United States) Vol. NS-29:1; ISSN IETNA
Country of Publication:
United States
Language:
English

Similar Records

An evaluation of the CRNL/CFFTP HT/HTO discriminating monitor
Journal Article · Thu Sep 01 00:00:00 EDT 1988 · Fusion Technol.; (United States) · OSTI ID:6135298

A Passive Diffusion Sampler for HT- and HTO-in-Air
Journal Article · Fri Jul 15 00:00:00 EDT 2005 · Fusion Science and Technology · OSTI ID:20854173

HT/HTO conversion in mammals
Conference · Sun Sep 01 00:00:00 EDT 1985 · Fusion Technol.; (United States) · OSTI ID:5804976