In-depth survey report: control technology for ethylene oxide sterilization at Community Midcenter Hospital, Marion, Ohio

Technical Report ·
OSTI ID:6073095

An in-depth survey of control technology for reducing exposure during ethylene oxide sterilization was conducted at Community Medcenter Hospital (SIC-8062), Marion, Ohio in October/November, 1984. Personal and general air monitoring for ethylene oxide was performed. Ventilation measurements were made. Engineering controls consisted of local exhaust ventilation, sterilizer cycle modification, and work practices. The authors conclude that the control system during the sterilizing operation provides adequate protection for the workers. Peak exposures during the time required to transfer the load may exceed the NIOSH recommended ceiling limit of 5ppm. Installation of better drain controls and improved exhaust ventilation of the recess room are recommended.