Performance of a PET detector module utilizing an array of silicon photodiodes to identify the crystal of interaction
Journal Article
·
· IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
- Lawrence Berkeley Lab., CA (United States)
- CTI, Inc., Knoxville, TN (United States)
The authors initial performance results for a new multi-layer PET detector module consisting of an array of 3 mm square by 30 mm deep BGO crystals coupled on one end to a single photomultiplier tube and on the opposite end to an array of 3 mm square silicon photodiodes. The photomultiplier tube provides an accurate timing pulse and energy discrimination for all the crystals in the module, while the silicon photodiodes identify the crystal of interaction. When a single BGO crystal at +25 C is excited with 511 keV photons, the authors measure a photodiode signal centered at 700 electrons (e[sup [minus]]) with noise of 375 e[sup [minus]] fwhm. When a four crystal/photodiode module is excited with a collimated line source of 511 keV photons, the crystal of interaction is correctly identified 82% of the time. The misidentification rate can be greatly reduced and an 8 x 8 crystal/photodiode module constructed by using thicker depletion layer photodiodes or cooling to 0 C.
- DOE Contract Number:
- AC03-76SF00098
- OSTI ID:
- 6062101
- Journal Information:
- IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States), Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) Vol. 40:4 part 1; ISSN 0018-9499; ISSN IETNAE
- Country of Publication:
- United States
- Language:
- English
Similar Records
Performance of a PET detector module utilizing an array of silicon photodiodes to identify the crystal of interaction
Performance of a PET detector module utilizing an array of silicon photodiodes to identify the crystal of interaction
Design studies for a PET detector module using a PIN photodiode to measure depth of interaction
Conference
·
Sat Oct 31 23:00:00 EST 1992
·
OSTI ID:6879193
Performance of a PET detector module utilizing an array of silicon photodiodes to identify the crystal of interaction
Conference
·
Sat Oct 31 23:00:00 EST 1992
·
OSTI ID:10138513
Design studies for a PET detector module using a PIN photodiode to measure depth of interaction
Conference
·
Mon Aug 01 00:00:00 EDT 1994
· IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
·
OSTI ID:6810831
Related Subjects
550601* -- Medicine-- Unsealed Radionuclides in Diagnostics
62 RADIOLOGY AND NUCLEAR MEDICINE
ACCURACY
COMPUTERIZED TOMOGRAPHY
DIAGNOSTIC TECHNIQUES
EMISSION COMPUTED TOMOGRAPHY
IMAGE SCANNERS
PERFORMANCE TESTING
PHOTODIODES
POSITRON COMPUTED TOMOGRAPHY
SEMICONDUCTOR DEVICES
SEMICONDUCTOR DIODES
TESTING
TOMOGRAPHY
62 RADIOLOGY AND NUCLEAR MEDICINE
ACCURACY
COMPUTERIZED TOMOGRAPHY
DIAGNOSTIC TECHNIQUES
EMISSION COMPUTED TOMOGRAPHY
IMAGE SCANNERS
PERFORMANCE TESTING
PHOTODIODES
POSITRON COMPUTED TOMOGRAPHY
SEMICONDUCTOR DEVICES
SEMICONDUCTOR DIODES
TESTING
TOMOGRAPHY