A personal monitoring method for SO/sub 2/ and NO/sub 2/ in the 1 to 10 ppM range, using triethanolamine (TEA) impregnated molecular sieve 13X as solid sorbent for sample collection was developed. Known concentration standards of SO/sub 2/ and NO/sub 2/ were generated by using liquid permeation tubes in a Kin-Tek three-channel gas calibration system. SO/sub 2/ and NO/sub 2/ were stripped from the sorbent with a standard eluant mixture (0.003 M NaHCO/sub 3//0.0024 M Na/sub 2/CO/sub 3/) and determined by ion chromatography as nitrite and/or nitrate and sulfate after oxidation of the sulfite with H/sub 2/O/sub 2/. Preparation and conditioning procedures for TEA-molecular sieve 13X, sample stability on the sorbent during storage, interferences of other species, effect of humidity on the recovery, and specific advantages of the proposed procedure over the existing methods are discussed.

Research Organization:
Monsanto Co., St. Louis, MO
OSTI ID:
6041545
Journal Information:
Anal. Chem.; (United States), Vol. 53:11
Country of Publication:
United States
Language:
English

Similar Records

Solid sorbent personal sampling method for the simultaneous collection of nitrogen dioxide and nitric oxide in air
Journal Article · Mon Aug 01 00:00:00 EDT 1977 · Am. Ind. Hyg. Assoc. J.; (United States) · OSTI ID:6041545

Ion chromatographic determination of thiosulfate in oil shale leachates
Journal Article · Thu Oct 01 00:00:00 EDT 1981 · Anal. Chem.; (United States) · OSTI ID:6041545

Explanation of pressure effects on a nitrogen dioxide (NO/sub 2/) sampler
Journal Article · Thu Jan 01 00:00:00 EST 1987 · Am. Ind. Hyg. Assoc. J.; (United States) · OSTI ID:6041545