We assessed particle deposition and loss occurring in two alpha-air monitors: an Eberline Alpha-3 Continuous Air Monitor (CAM) and a working-area transuranic aerosol monitor (WOTAMS). We investigated the dependence of particle size on losses in the sampling inlets and the real-time alpha detector areas for both instruments. We determined the uniformity of particle deposition on the filter to ascertain the effectiveness of the detector and collection-filter configuration. Results indicate that particle losses are a strong function of particle size in the CAM unit, with a 44% loss occurring for 6-..mu..m-diameter aerosols and a 0.3% loss for 0.6-..mu..m-diameter aerosols. Losses in the WOTAMS were less than 1% for particle diameters in the 0.6-to-7 ..mu..m range.
Research Organization:
Lawrence Livermore National Lab., CA (USA)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
6040298
Report Number(s):
UCRL-93521; CONF-8510235-3; ON: DE86006706
Country of Publication:
United States
Language:
English