Influence of the sample position in compensated scanning calorimeters
Journal Article
·
· Anal. Chem.; (United States)
In determining the influence of sample position on the measured signal in compensated scanning calorimeters, Swedish researchers from the University of Lund found that the signal measured is equal to a thermal signal generated in the sample only under the conditions that 1) the investigated sample is inside the compensation feedback loop and 2) the gain of the amplifier is sufficiently high. The theoretical analysis revealed that when the sample is outside the feedback loop, the signal measured depends also on the time constant of the sample and the heating rate, even if the gain of the amplifier is very high.
- OSTI ID:
- 6038908
- Journal Information:
- Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 50; ISSN ANCHA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Sensitive power compensated scanning calorimeter for analysis of phase transformations in small samples
Log amplifier with pole-zero compensation
Log amplifier with pole-zero compensation
Journal Article
·
Wed Jun 15 00:00:00 EDT 2005
· Review of Scientific Instruments
·
OSTI ID:20722919
Log amplifier with pole-zero compensation
Patent
·
Wed Dec 31 23:00:00 EST 1986
·
OSTI ID:866105
Log amplifier with pole-zero compensation
Patent
·
Thu Feb 07 23:00:00 EST 1985
·
OSTI ID:5271248