Characterization of new focus projection and scan (FPS) vidicons for scientific imaging applications

Conference ·
DOI:https://doi.org/10.1117/12.942244· OSTI ID:6029035
Several new photoconductors now commercially available as targets in Type 7803 Focus Projection and Scan (FPS) electrostatically focused vidicons have been characterized for use as radiometric sensors in transient illumination and single frame applications. These include Saticon (Se + As + Te), Newvicon (ZnSe), Pasecon (CdSe), and Plumbicon (PbO). Samples from several domestic and foreign manufacturers have been evaluated for photoconductive response time and responsivity at selected narrow wavelength bands, including 410 nm, 560 nm, and 822 nm. These data are compared with performance data from older target materials including Sb/sub 2/S/sub 3/ and silicon. The effect of bias lighting on sensitivity and transfer curve slopes for single-event pulsed light stimulas are presented. Dynamic range and saturation limits as functions of beam aperture, target voltage, and filament current are also discussed. 6 refs., 17 figs., 3 tabs.
Research Organization:
Los Alamos National Lab., NM (USA)
DOE Contract Number:
W-7405-ENG-36
OSTI ID:
6029035
Report Number(s):
LA-UR-87-2712; CONF-8708110-10; ON: DE87014728
Country of Publication:
United States
Language:
English