Peroxyoxalate chemiluminescence detection of polycyclic aromatic amines in liquid chromatograhpy

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00271a011· OSTI ID:6011661
Subpicogram detection limits are obtained for many aminosubstituted polycyclic aromatic hydrocarbons (amino-PAH) in high-performance liquid chromatography (HPLC) using peroxyoxalate chemiluminescence detection. Detection limits are typically 1 to 2 orders of magnitude lower than those that can be achieved in the same chromatographic system using fluorescence detection. The high selectivity of the chemiluminescence detection for amino-PAH allows these compounds to be determined in complex matrices without complete chromatographic separation. The HPLC/chemiluminescence method is applied to the detection of amino-PAH in samples of shale oil, coal oil, and coal gasifier tar and the results are compared with fused silica capillary column gas chromatography coupled with various means of detection, including mass spectrometry. 32 references, 9 figures, 4 tables.
Research Organization:
Univ. of Colorado, Boulder
OSTI ID:
6011661
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 56:7; ISSN ANCHA
Country of Publication:
United States
Language:
English