Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Characterization of a large-format, high-fidelity, picosecond, optical streak camera

Conference ·
OSTI ID:6004975
We have completed preliminary characterizations of a new 3-grid, large-format, ITT streak camera tube operating with about a 2.3-kV extraction voltage. We have found that the best attainable impulse response is approx.10 ps at low input intensity. However, the impulse response increases to only 30 ps with an increase of nearly three orders of magnitude in input intensity. Further, the spatial resolution remains greater that 5 line pairs/mm at 50% contrast over these three orders of magnitude of input intensity. We have also done preliminary computational modeling of the tube's temporal response as a function of focus voltage, illumination spot size, and wavelength. 8 refs.
Research Organization:
Lawrence Livermore National Lab., CA (USA); Test and Measurement Systems, Inc., Santa Clara, CA (USA); Compagnie Generale de Telegraphie Sans Fil (CSF), 75 - Paris (France)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
6004975
Report Number(s):
UCRL-96920; CONF-8708110-25; ON: DE88001594
Country of Publication:
United States
Language:
English