Intracavity absorption with external fluorescence measurement for detection of radioiodine isotopes

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac50043a034· OSTI ID:5992941

We report the use of intracavity absorption in a CW dye laser as a sensitive analytical method for the detection of iodine-129. With an external fluorescence detection scheme employing photon counting, quantitative isotope specific measurements of /sup 127/I/sub 2/ and /sup 129/I/sub 2/ have been made at concentrations in the range of 5 x 10/sup 11/ to 7 x 10/sup 15/ molecules/cm/sup 3/. Scaling estimates indicate that a lower level may be achieved if desirable. In addition, this method permits the simultaneous detection of a number of iodine isotopes.

Research Organization:
Sandia Labs., Albuquerque, NM
OSTI ID:
5992941
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 51:7; ISSN ANCHA