Performance study of whole-body, multislice positron computed tomograph: POSITOLOGICA-II
A 3-detector ring, 5-slice whole-body positron CT has been developed and is being tested. The PCT, POSITOLOGICA-II, has a total of 480 BGO's (160/ring), and employs continuous rotation scan (0.5rps). By using a 15mm wide BGO, a 9.2mm FWHM spatial resolution for reconstructed image is obtained at the center of FOV. Measured phantom diameter dependence of the true count rate shows good agreement with theoretically anticipated characteristics, including maximum sensitivity at around 30cm dia. Sensitivities for 20cm dia. phantom are 28 and 38 kcps/..mu..Ci/ml for in-plane and cross-plane, respectively, including scattered coincidences. Relative system detection efficiency measured with a line source at FOV center is 96% at 15ns time window (90% at 12ns), basing on 100% efficiency in 2024ns.
- Research Organization:
- Hitachi Central Research Laboratory, Kokubunji, Tokyo
- OSTI ID:
- 5990943
- Journal Information:
- IEEE Trans. Nucl. Sci.; (United States), Vol. 30:1
- Country of Publication:
- United States
- Language:
- English
Similar Records
HEADTOME III: A high quantitation and high resolution brain positron emission tomograph
A high resolution animal PET scanner using compact PS-PMT detectors
Related Subjects
62 RADIOLOGY AND NUCLEAR MEDICINE
POSITRON COMPUTED TOMOGRAPHY
PERFORMANCE
RADIATION DETECTORS
PERFORMANCE TESTING
BISMUTH COMPOUNDS
EFFICIENCY
GERMANIUM OXIDES
IMAGES
PHANTOMS
SCATTERING
SENSITIVITY
SPATIAL RESOLUTION
CHALCOGENIDES
COMPUTERIZED TOMOGRAPHY
DIAGNOSTIC TECHNIQUES
EMISSION COMPUTED TOMOGRAPHY
GERMANIUM COMPOUNDS
MEASURING INSTRUMENTS
MOCKUP
OXIDES
OXYGEN COMPOUNDS
RESOLUTION
STRUCTURAL MODELS
TESTING
TOMOGRAPHY
440100* - Radiation Instrumentation
550600 - Medicine