Effect of increased axial field of view on the performance of a volume PET scanner
- Univ. of Pennsylvania, Philadelphia, PA (United States). Dept. of Radiology
- UGM Medical Systems Inc., Philadelphia, PA (United States)
The performance of the PENN-PET 240H scanner from UGM Medical Systems is tested and compared to the prototype PENN-PET scanner built at the University of Pennsylvania. The UGM PENN-PET scanner consists of six continuous position-sensitive NaI(Tl) detectors, which results in a 50 cm transverse field-of-view and a 12.8 cm axial field-of-view. The fine spatial sampling in the axial direction allows the data to be sorted into as many as 64 transverse planes, each 2 mm thick. A large axial acceptance angle, without interplane septa, results in a high-sensitivity and low-randoms fraction, with a low-scatter fraction due to the use of a narrow photopeak energy window. This paper emphasizes those performance measurements that illustrate the special characteristics of a volume imaging scanner and how they change as the axial length is increased.
- DOE Contract Number:
- FG02-88ER60642
- OSTI ID:
- 5953379
- Journal Information:
- IEEE Transactions on Medical Imaging (Institute of Electrical and Electronics Engineers); (United States), Journal Name: IEEE Transactions on Medical Imaging (Institute of Electrical and Electronics Engineers); (United States) Vol. 12:2; ISSN 0278-0062; ISSN ITMID4
- Country of Publication:
- United States
- Language:
- English
Similar Records
Effect of increased axial field of view of on the performance of a volume PET scanner. [Positron Emission Tomography (PET)]
Design and performance of HEAD PENN-PET scanner
Related Subjects
62 RADIOLOGY AND NUCLEAR MEDICINE
ACCURACY
CHARGED PARTICLE DETECTION
COMPUTERIZED TOMOGRAPHY
DETECTION
DIAGNOSTIC TECHNIQUES
EMISSION COMPUTED TOMOGRAPHY
IMAGE SCANNERS
PERFORMANCE TESTING
POSITRON COMPUTED TOMOGRAPHY
POSITRON DETECTION
RADIATION DETECTION
TESTING
TOMOGRAPHY