Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Dose parameters of /sup 125/I and /sup 192/Ir seed sources

Journal Article · · Med. Phys.; (United States)
OSTI ID:5937387
As mandated by an NCI brachytherapy contract, we measured dosimetric parameters for /sup 192/Ir seeds and two models of /sup 125/I seeds. Measurements were with LiF powder in a water-equivalent phantom. Data were corrected for background, sample mass, and finite detector volume. Selected parameters were also investigated through Monte Carlo calculations. Results are presented in terms of a dose parametrization that is described in detail, and are compared to published data. Our results agreed well with published data for relative quantities such as radial and angular dose dependence. Our measured value for the /sup 192/Ir dose factor was 4.55 cGy(H/sub 2/O) cm/sup 2/ mCi/sup /minus/1/ h/sup /minus/1/, also in good agreement with commonly used values. However, the measured dose factors for /sup 125/I seed models 6702 and 6711 were 1.18 and 1.06 cGy(H/sub 2/O) cm/sup 2/ mCi/sup /minus/1/ h/sup /minus/1/, values well below those in general use.
Research Organization:
Department of Radiation Oncology, University of California, San Francisco, California 94143(US)
OSTI ID:
5937387
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 16:4; ISSN MPHYA
Country of Publication:
United States
Language:
English