Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Comparative measurements of potassium and chloride with ion-sensitive microelectrodes and x-ray microanalysis in cultured skeletal muscle fibers

Journal Article · · In Vitro Cell. Develop. Biol.; (United States)
DOI:https://doi.org/10.1007/BF02620913· OSTI ID:5850267
Data of the intracellular electrolyte concentration of potassium and chloride in cultured muscle cells measured by x-ray analysis were compared by using the different activity coefficients with intracellular potassium and chloride activities measured with double-barrelled microelectrodes. By using an activity coefficient of 0.6, 95% of the potassium microelectrode measurements are in accordance with the x-ray analysis values, in spite of a scattering of the values. Membrane potential and intracellular potassium values are linearly related. x-ray analysis and ion-sensitive microelectrodes measured the cytoplasmic chloride in the same range. Taking into account known activity coefficients, an error of 25% must be assumed with the intracellular chloride measurements. However, x-ray analysis and ion-sensitive microelectrode investigations are reliable tools to study intracellular potassium and chloride changes, which play an important role in membrane characteristics.
Research Organization:
Max-Planck Institut fuer Systemphysiologie, Dortmund, Germany, F.R.
OSTI ID:
5850267
Journal Information:
In Vitro Cell. Develop. Biol.; (United States), Journal Name: In Vitro Cell. Develop. Biol.; (United States) Vol. 1; ISSN ICDBE
Country of Publication:
United States
Language:
English