Design and properties of a multianode, cylindrical proportional counter for position sensing at high count rates

Journal Article · · Rev. Sci. Instrum.; (United States)
DOI:https://doi.org/10.1063/1.1135677· OSTI ID:5837617
A cylindrical proportional counter with six anode wires was designed and tested. The electron drift velocity was greater than that of similar single-anode, coaxial proportional counter, thus, allowing counting and position sensing of low-energy photons (<60 keV) at high count rates (>10/sup 5/ counts/s).
Research Organization:
Department of Radiology, University of Wisconsin Hospitals, Madison, Wisconsin 53706
OSTI ID:
5837617
Journal Information:
Rev. Sci. Instrum.; (United States), Journal Name: Rev. Sci. Instrum.; (United States) Vol. 50:10; ISSN RSINA
Country of Publication:
United States
Language:
English