Design and properties of a multianode, cylindrical proportional counter for position sensing at high count rates
Journal Article
·
· Rev. Sci. Instrum.; (United States)
A cylindrical proportional counter with six anode wires was designed and tested. The electron drift velocity was greater than that of similar single-anode, coaxial proportional counter, thus, allowing counting and position sensing of low-energy photons (<60 keV) at high count rates (>10/sup 5/ counts/s).
- Research Organization:
- Department of Radiology, University of Wisconsin Hospitals, Madison, Wisconsin 53706
- OSTI ID:
- 5837617
- Journal Information:
- Rev. Sci. Instrum.; (United States), Journal Name: Rev. Sci. Instrum.; (United States) Vol. 50:10; ISSN RSINA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Multianode cylindrical proportional counter for high count rates
Multianode cylindrical proportional counter for high count rates
Multianode cylindrical proportional counter for high count rates
Patent
·
Tue Sep 15 00:00:00 EDT 1981
·
OSTI ID:6305546
Multianode cylindrical proportional counter for high count rates
Patent
·
Wed Dec 31 23:00:00 EST 1980
·
OSTI ID:863986
Multianode cylindrical proportional counter for high count rates
Patent
·
Fri May 23 00:00:00 EDT 1980
·
OSTI ID:6604127
Related Subjects
440101* -- Radiation Instrumentation-- General Detectors or Monitors & Radiometric Instruments
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
CONFIGURATION
CYLINDRICAL CONFIGURATION
DESIGN
MEASURING INSTRUMENTS
MULTIWIRE PROPORTIONAL CHAMBERS
PERFORMANCE TESTING
POSITION SENSITIVE DETECTORS
PROPORTIONAL COUNTERS
RADIATION DETECTORS
TESTING
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
CONFIGURATION
CYLINDRICAL CONFIGURATION
DESIGN
MEASURING INSTRUMENTS
MULTIWIRE PROPORTIONAL CHAMBERS
PERFORMANCE TESTING
POSITION SENSITIVE DETECTORS
PROPORTIONAL COUNTERS
RADIATION DETECTORS
TESTING