skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Low background Ge(Li) detector gamma-ray spectroscopy system with sample changer

Technical Report ·
DOI:https://doi.org/10.2172/5800280· OSTI ID:5800280

A gamma-ray spectroscopy system used for analysis of volume soil and vegetation samples is described. It can accept samples of up to 2 kg weight and a 970 cm/sup 3/ volume. Its primary use is in the determination of U, Th, /sup 137/Cs, /sup 40/K, and /sup 60/Co in samples obtained during ground truth AMS surveys throughout the country. The system was calibrated using standards from the DOE New Brunswick Laboratory and from the National Bureau of Standards. Detectabilities (at 3sigma) in a 1500-g sample are: 0.28 ppM Th, 0.10 ppM U, 10 pCi /sup 40/K, 1.0 pCi ./sup 137/Cs, and 1.0 pCi of /sup 60/Co.

Research Organization:
EG and G, Inc., Goleta, CA (USA). Energy Measurements Group
DOE Contract Number:
EY-76-C-08-1183
OSTI ID:
5800280
Report Number(s):
EGG-1183-2383; TRN: 80-002618
Country of Publication:
United States
Language:
English