Potentiostatic coulometric determination of iridium with the use of a glassy carbon electrode

Journal Article · · J. Anal. Chem. USSR (Engl. Transl.); (United States)
OSTI ID:5785341

An industrially produced glassy carbon crucible has been used for the potentiostatic coulometric determination of iridium as an electrochemical cell, a working electrode, and a vessel for weighing an aliquot portion of the solution. A method for the coulometric determination of 1-7 mg of iridium in a 0.5 M HCl supporting electrolyte solution has been developed. The value of s/sur r/ is 0.002 to 0.008. The results of the determination of iridium in solutions of industrial wastes are characterized by values of s/sub r/ ranging from 0.002 to 0.006.

Research Organization:
V. I. Vernadskii Institute of Geochemistry and Analytical Chemistry, Moscow, USSR
OSTI ID:
5785341
Journal Information:
J. Anal. Chem. USSR (Engl. Transl.); (United States), Journal Name: J. Anal. Chem. USSR (Engl. Transl.); (United States) Vol. 41:1; ISSN JACUA