This paper reports on a prototype skin dose monitor that has been constructed and tested successfully. The monitor uses an ion-implanted silicon detector available commercially. The ion-implanted silicon detector was selected since it has a lower leakage current than the other type of silicon detectors. It operates on both a pulse mode and a current mode, and there is an overlap in the measurable beta dose rate ranges of these two modes. Our prototype device includes an 8-bit A/D converter and an 8-bit microprocessor to display the skin dose directly. The device presently covers more than five orders of magnitude in the measurable beta skin dose rate ranges. The device has been calibrated using an extrapolation chamber and is then used to measure the dose rates produced by several different beta and beta/gamma sources.
- OSTI ID:
- 5760020
- Journal Information:
- IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States), Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) Vol. 38:4; ISSN 0018-9499; ISSN IETNA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Research and development of a beta skin-dose monitor using silicon detectors
Skin dose measurement with MICROSPEC-2{trademark}
Microcomputer Monte Carlo electron transport codes for beta skin dose calculations
Thesis/Dissertation
·
Mon Dec 31 23:00:00 EST 1990
·
OSTI ID:6913558
Skin dose measurement with MICROSPEC-2{trademark}
Technical Report
·
Wed Oct 01 00:00:00 EDT 1997
·
OSTI ID:532502
Microcomputer Monte Carlo electron transport codes for beta skin dose calculations
Journal Article
·
Sat Jun 01 00:00:00 EDT 1991
· IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
·
OSTI ID:5680019
Related Subjects
440101* -- Radiation Instrumentation-- General Detectors or Monitors & Radiometric Instruments
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
BETA DOSIMETRY
BETA SOURCES
CALIBRATION
DOSE RATES
DOSIMETRY
EXTRAPOLATION
FABRICATION
GAMMA SOURCES
ION IMPLANTATION
MEASURING INSTRUMENTS
MONITORS
NUMERICAL SOLUTION
PARTICLE SOURCES
PERFORMANCE TESTING
RADIATION DETECTORS
RADIATION MONITORS
RADIATION SOURCES
SEMICONDUCTOR DETECTORS
SI SEMICONDUCTOR DETECTORS
TESTING
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
BETA DOSIMETRY
BETA SOURCES
CALIBRATION
DOSE RATES
DOSIMETRY
EXTRAPOLATION
FABRICATION
GAMMA SOURCES
ION IMPLANTATION
MEASURING INSTRUMENTS
MONITORS
NUMERICAL SOLUTION
PARTICLE SOURCES
PERFORMANCE TESTING
RADIATION DETECTORS
RADIATION MONITORS
RADIATION SOURCES
SEMICONDUCTOR DETECTORS
SI SEMICONDUCTOR DETECTORS
TESTING