Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Comparison of neutron/gamma dose estimates: ion-chamber measurements vs. Monte Carlo calculations

Technical Report ·
OSTI ID:5733549
Ion-chamber measurements in a water phantom irradiated by neutrons from the Gas Target Neutron Source have been completed at depths of 3.5, 7.5, 10, 15 and 20 cm. It is now possible to construct a three-dimensional dose profile for both neutron- and gamma-absorbed doses. A novel way of correcting for the displacement of water by the ion chambers and GM counter was utilized. Comparison of these results with previous Monte Carlo calculations indicates problems with neutrons scattered within the irradiation room.
Research Organization:
Wisconsin Univ., Madison (USA). Dept. of Medical Physics
DOE Contract Number:
AC02-76EV01105
OSTI ID:
5733549
Report Number(s):
DOE/EV/01105-303; ON: DE83017601
Country of Publication:
United States
Language:
English