Report on the test and evaluation of the Chaparral Physics Model 4.1.1 prototype microbarograph for CTBT infrasound array application

Technical Report ·
DOI:https://doi.org/10.2172/573309· OSTI ID:573309
 [1];  [2]
  1. Sandia National Labs., Albuquerque, NM (United States). Monitoring Technologies Dept.
  2. Applied Physics Inc., Albuquerque, NM (United States)
The Sandia National Laboratories has tested and evaluated the Chaparral Physics Model 4.1.1 prototype infrasound sensor to CTBT specifications. The sensor was characterized by using a piston-phone chamber to set and measure sensor sensitivity. Multiple sensor side-by-side coherence analysis testing provided a measure of sensor relative gain and phase; sensor self-noise was computed using this technique. The performance of the sensor calibration circuitry was evaluated. Sensor performance was compared to CTBT specifications. The Chaparral sensor met or exceeded all CTBT specifications.
Research Organization:
Sandia National Labs., Albuquerque, NM (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC04-94AL85000
OSTI ID:
573309
Report Number(s):
SAND--98-0127; ON: DE98002850
Country of Publication:
United States
Language:
English

Similar Records

2022 Chaparral 64S Infrasound Sensor Type Approval Evaluation
Technical Report · Fri Jul 01 00:00:00 EDT 2022 · OSTI ID:1874854

Test report for the infrasound prototype: For a CTBT IMS station
Technical Report · Fri Oct 31 23:00:00 EST 1997 · OSTI ID:552766

Chaparral Model 60 Infrasound Sensor Evaluation.
Technical Report · Mon Feb 29 23:00:00 EST 2016 · OSTI ID:1431260