Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

High frame-rate neutron radiography of dynamic events

Conference ·
OSTI ID:5685357
A system has been developed to perform neutron radiographic analysis of dynamic events having a duration of several milliseconds. The system has been operated in the range of 2000 to 10,000 frames/second. Synchronization has provided high-speed-motion neutron radiographs for evaluation of the firing cycle of 7.62 mm munition rounds within a steel rifle barrel. The system has also been used to demonstrate the ability to produce neutron radiographic movies of two-phase flow. The equipment uses the Oregon State University TRIGA reactor capable of pulsing to 3000 MW peak power, a neutron beam collimator, a scintillator neutron conversion screen coupled to an image intensifier, and a 16 mm high speed movie camera. The peak neutron flux incident at the object position is approximately 4 x 10/sup 11/ n/cm/sup 2/s with a pulse, full width at half maximum, of 9 ms. Special studies have been performed on the scintillator conversion screens and on the effects of statistical limitations on the image quality. Modulation transfer function analysis has been used to assist in the evaluation of the system performance.
Research Organization:
Lawrence Livermore National Lab., CA (USA); Oregon State Univ., Corvallis (USA)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
5685357
Report Number(s):
UCRL-86954; CONF-811215-1; ON: DE82004389
Country of Publication:
United States
Language:
English