A gamma-ray converter for a multiwire proportional chamber (MWPC) positron emission tomograph is described. The converter is made of small-diameter (0.48 mm inner diameter, 0.06 mm wall thickness) lead-oxide-glass tubes fused to form a honeycomb matrix. The surfaces of the tubes are reduced in a hydrogen atmosphere to provide the drift electric field for detection of the conversion electrons. The detection efficiency for a 10 mm thick converter is 6.65%, with a time resolution of 160 ns (FWHM). A scheme which will improve the spatial resolution of the tomograph by use of the self quenching streamer mode of chamber operation is described. Details of construction of the converters and the MWPC are presented, as well as the design performance of a high spatial resolution positron emission tomograph (HISPET). 40 refs., 22 figs.