Operating manual for the electrostatic glove-box prefilter installed inside the filter glove box No. 046 at Rocky Flats, Building 776
Objective of the evaluation is to evaluate the effectiveness of the electrostatic prefilter in prolonging the life of HEPA (high-efficiency particulate-air) filters. The theory of the electrostatic filter is reviewed, and Glove Box Number 046 is described in detail, followed by a description of the electrostatic prefilter used in the present application. Engineering drawings of the electrostatic prefilter are included. The procedure for evaluating the electrostatic prefilter includes the steps for conducting five different tests: evaluating (1) the HEPA filter alone, (2 and 3) the HEPA filter with a standard prefilter treated both as disposable and reusable, and (4 and 5) the HEPA filter with the electrostatic prefilter, again treated as disposable and reusable. Procedures for flowmeter calibrations and measurements of particle-size distributions are also included. Long-term maintenence of the system during the evaluation program is outlined, and estimates of component durability are given. An electrical engineering safety note describes the high-voltage operational hazard of the electrostatic prefilter and the testing of safety devices (the current-overload trip circuit, the filter-door-interlock system, and the current-limiting resistor).
- Research Organization:
- California Univ., Livermore (USA). Lawrence Livermore Lab.; Atomics International Div., Golden, CO (USA). Rocky Flats Plant
- DOE Contract Number:
- W-7405-ENG-48
- OSTI ID:
- 5677691
- Report Number(s):
- UCID-18180
- Country of Publication:
- United States
- Language:
- English
Similar Records
Electrofibrous prefilters for use in the nuclear industry
Evaluation of prototype electrofibrous filters for nuclear-ventilation ducts
Electrofibrous prefilters for use in nuclear ventilation systems
Conference
·
Mon Dec 31 23:00:00 EST 1979
·
OSTI ID:5279234
Evaluation of prototype electrofibrous filters for nuclear-ventilation ducts
Conference
·
Wed Dec 01 23:00:00 EST 1982
·
OSTI ID:6644488
Electrofibrous prefilters for use in nuclear ventilation systems
Conference
·
Wed Feb 18 23:00:00 EST 1981
·
OSTI ID:6367795
Related Subjects
42 ENGINEERING
420203* -- Engineering-- Handling Equipment & Procedures
AIR FILTERS
CALIBRATION
ELECTROSTATIC PRECIPITATORS
EQUIPMENT
EVALUATION
FILTERS
FLOWMETERS
GLOVEBOXES
LABORATORY EQUIPMENT
MAINTENANCE
MEASURING INSTRUMENTS
PARTICLE SIZE
PERFORMANCE TESTING
POLLUTION CONTROL EQUIPMENT
SAFETY
SIZE
TESTING
420203* -- Engineering-- Handling Equipment & Procedures
AIR FILTERS
CALIBRATION
ELECTROSTATIC PRECIPITATORS
EQUIPMENT
EVALUATION
FILTERS
FLOWMETERS
GLOVEBOXES
LABORATORY EQUIPMENT
MAINTENANCE
MEASURING INSTRUMENTS
PARTICLE SIZE
PERFORMANCE TESTING
POLLUTION CONTROL EQUIPMENT
SAFETY
SIZE
TESTING