Computer programs for output and depth dose from hyperbolic equations

Journal Article · · Med. Phys.; (United States)
DOI:https://doi.org/10.1118/1.594916· OSTI ID:5628506
A compact and accurate equation has been developed that describes the output field size over the range from 4 x 4 to 35 x 35 cm2. An accuracy of +/- 0.2% has been demonstated for teletherapy machines with energies from 250 kVp to 6 MV. The same accuracy down to a depth of at least 30 cm. The relationship is described by a hyperbola whose x and y axes are displaced by corresponding values X0 and Y0 producing the equation y = Y0 - C/(X0 + x), where y = output and x = field size. By substituting three separate values for y and x into three simultaneous equations, values for Y0, X0 and c can be derived.