TFTR Michelson interferometer electron cyclotron emission diagnostic

Journal Article · · Rev. Sci. Instrum.; (United States)
DOI:https://doi.org/10.1063/1.1138044· OSTI ID:5594510
In July 1984, a Fourier transform spectrometer employing a fast-scanning Michelson interferometer began operating on TFTR. This diagnostic system can measure the electron cyclotron emission spectrum 72 times per s with a time resolution of 11 ms and a spectral resolution of 3.6 GHz. The initial operating spectral range is 75--540 GHz, which is adequate for measuring the first three cyclotron harmonics at present TFTR magnetic field levels. The range can be extended easily to 75--1080 GHz in order to accommodate increases in toroidal magnetic field or to study superthermal ECE. The measured spectra are absolutely calibrated using a liquid nitrogen cooled blackbody reference source. The second harmonic feature of each spectrum is used to calculate the absolute electron temperature profile.
Research Organization:
Laboratory for Plasma and Fusion Energy Studies, University of Maryland, College Park, Maryland 20742
OSTI ID:
5594510
Journal Information:
Rev. Sci. Instrum.; (United States), Journal Name: Rev. Sci. Instrum.; (United States) Vol. 56:5; ISSN RSINA
Country of Publication:
United States
Language:
English