Low-frequency nuclear magnetic resonance and nuclear quadrupole resonance spectrometer based on a dc superconducting quantum interference device
Journal Article
·
· Review of Scientific Instruments; (USA)
- Department of Physics, University of California and Materials Sciences Division, Lawrence Berkeley Laboratory, Berkeley, California 94720 (USA)
A sensitive spectrometer, based on a dc superconducting quantum interference device, for the direct detection of low-frequency pulsed nuclear magnetic resonance (NMR) and nuclear quadrupole resonance (NQR), is described. The frequency response extends from about 10 to 200 kHz, and the recovery time after the magnetic pulse is removed is typically 50 {mu}s. As examples, NMR spectra are shown from Pt and Cu metal powders in a magnetic field of 6 mT, and NQR spectra are shown from {sup 2}D in a tunneling methyl group and {sup 14}N in NH{sub 4}ClO{sub 4}.
- DOE Contract Number:
- AC03-76SF00098
- OSTI ID:
- 5550680
- Journal Information:
- Review of Scientific Instruments; (USA), Journal Name: Review of Scientific Instruments; (USA) Vol. 62:6; ISSN RSINA; ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Low-Frequency NMR and NQR with a dc SQUID Amplifier [Thesis]
Direct current superconducting quantum interference device spectrometer for pulsed nuclear magnetic resonance and nuclear quadrupole resonance at frequencies up to 5 MHz
Nuclear magnetic resonance with dc SQUID (Super-conducting QUantum Interference Device) preamplifiers
Thesis/Dissertation
·
Wed Oct 31 23:00:00 EST 1990
·
OSTI ID:5775965
Direct current superconducting quantum interference device spectrometer for pulsed nuclear magnetic resonance and nuclear quadrupole resonance at frequencies up to 5 MHz
Journal Article
·
Thu Aug 01 00:00:00 EDT 1996
· Review of Scientific Instruments
·
OSTI ID:286880
Nuclear magnetic resonance with dc SQUID (Super-conducting QUantum Interference Device) preamplifiers
Conference
·
Mon Aug 01 00:00:00 EDT 1988
·
OSTI ID:6225439
Related Subjects
440800* -- Miscellaneous Instrumentation-- (1990-)
47 OTHER INSTRUMENTATION
656002 -- Condensed Matter Physics-- General Techniques in Condensed Matter-- (1987-)
75 CONDENSED MATTER PHYSICS
SUPERCONDUCTIVITY AND SUPERFLUIDITY
DESIGN
ELECTRONIC EQUIPMENT
ELEMENTS
EQUIPMENT
FLUXMETERS
FREQUENCY RANGE
HZ RANGE
KHZ RANGE
MEASURING INSTRUMENTS
METALS
MICROWAVE EQUIPMENT
NMR SPECTROMETERS
NUCLEAR QUADRUPOLE RESONANCE
PLATINUM
PLATINUM METALS
RESONANCE
SPECTROMETERS
SQUID DEVICES
SUPERCONDUCTING DEVICES
TRANSITION ELEMENTS
USES
47 OTHER INSTRUMENTATION
656002 -- Condensed Matter Physics-- General Techniques in Condensed Matter-- (1987-)
75 CONDENSED MATTER PHYSICS
SUPERCONDUCTIVITY AND SUPERFLUIDITY
DESIGN
ELECTRONIC EQUIPMENT
ELEMENTS
EQUIPMENT
FLUXMETERS
FREQUENCY RANGE
HZ RANGE
KHZ RANGE
MEASURING INSTRUMENTS
METALS
MICROWAVE EQUIPMENT
NMR SPECTROMETERS
NUCLEAR QUADRUPOLE RESONANCE
PLATINUM
PLATINUM METALS
RESONANCE
SPECTROMETERS
SQUID DEVICES
SUPERCONDUCTING DEVICES
TRANSITION ELEMENTS
USES