Effects of neutron energy spectrum on the efficiency calibration of epithermal neutron detectors

Journal Article · · Rev. Sci. Instrum.; (United States)
DOI:https://doi.org/10.1063/1.1138230· OSTI ID:5545359
The /sup 235/U epithermal neutron fission detector system at TFTR has been calibrated in situ, using three different neutron sources: /sup 252/Cf, a D-D, and a D-T neutron generator. These sources produced a continuous spectrum, a line spectrum at approx.2.5 MeV, and a line spectrum at approx.14 MeV, respectively. The toroidally averaged detection efficiencies are 4.61 x 10/sup -9/ (/sup 252/Cf), 4.80 x 10/sup -9/ (D-D), and 5.42 x 10/sup -9/ (D-T) counts/neutron for a moderated 1.3-g /sup 235/U detector placed approx.4 m from the torus centerline. The effects of (n,2n) and (n,n'..gamma..) reactions are discussed.
Research Organization:
Princeton University, Plasma Physics Laboratory, Princeton, New Jersey 08544