Determination of hydrogen in uranium-niobium-zirconium alloy by inert-gas fusion

Technical Report ·
DOI:https://doi.org/10.2172/5529050· OSTI ID:5529050
An improved method has been developed using inert-gas fusion for determining the hydrogen content in uranium-niobium-zirconium (U-7.5Nb-2.5Zr) alloy. The method is applicable to concentrations of hydrogen ranging from 1 to 250 micrograms per gram and may be adjusted for analysis of greater hydrogen concentrations. Hydrogen is determined using a hydrogen determinator. The limit of error for a single determination at the 95%-confidence level (at the 3.7-..mu..g/g-hydrogen level) is +-1.4 micrograms per gram hydrogen.