The effects of nuclear magnetic resonance on patients with cardiac pacemakers
                            Journal Article
                            ·
                            
                            · Radiology; (United States)
                            
                        
                    - Cleveland Clinic Foundation, OH
The effect of nuclear magnetic resonance (NMR) imaging on six representative cardiac pacemakers was studied. The results indicate that the threshold for initiating the asynchronous mode of a pacemaker is 17 gauss. Radiofrequency levels are present in an NMR unit and may confuse or possibly inhibit demand pacemakers, although sensing circuitry is normally provided with electromagnetic interference discrimination. Time-varying magnetic fields can generate pulse amplitudes and frequencies to mimic cardiac activity. A serious limitation in the possibility of imaging a patient with a pacemaker would be the alteration of normal pulsing parameters due to time-varying magnetic fields.
- OSTI ID:
- 5504990
- Journal Information:
- Radiology; (United States), Journal Name: Radiology; (United States) Vol. 147:1; ISSN RADLA
- Country of Publication:
- United States
- Language:
- English
Similar Records
                                
                                
                                    
                                        
                                        Effect of 1. 5 tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers
                                        
Effects of 60-Heartz electric and magnetic fields on implanted cardiac pacemakers. Final report. [Hazards of power transmission line frequencies]
Proton Beam Therapy Interference With Implanted Cardiac Pacemakers
                        
                                            Journal Article
                                            ·
                                            Thu Oct 01 00:00:00 EDT 1987
                                            · J. Am. Coll. Cardiol.; (United States)
                                            ·
                                            OSTI ID:5851484
                                        
                                        
                                        
                                    
                                
                                    
                                        Effects of 60-Heartz electric and magnetic fields on implanted cardiac pacemakers. Final report. [Hazards of power transmission line frequencies]
                                            Technical Report
                                            ·
                                            Sat Sep 01 00:00:00 EDT 1979
                                            
                                            ·
                                            OSTI ID:5729932
                                        
                                        
                                        
                                    
                                
                                    
                                        Proton Beam Therapy Interference With Implanted Cardiac Pacemakers
                                            Journal Article
                                            ·
                                            Sat Nov 01 00:00:00 EDT 2008
                                            · International Journal of Radiation Oncology, Biology and Physics
                                            ·
                                            OSTI ID:21128202